Agilent Technologies 54657A Especificaciones Pagina 99

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 188
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 98
4 Press any key once to advance the test.
The multimeter should read 5.000 V ± 10 mV. If the result is not within the
test limits, see "Troubleshooting the oscilloscope," on page 4-32.
5
Connect an oscilloscope to the rear-panel CALIBRATOR connector.
6 Press any key once to advance the test.
7 Obtain a stable display on the oscilloscope.
8 Measure Vp--p, Vavg, and the frequency of the signal.
If the results are not within the Delay Calibrator test limits stated above, see
"Troubleshooting the oscilloscope," on page 4-32.
9
Press any key once to end this test.
Service
Verifying Oscilloscope Performance
4-7
Vista de pagina 98
1 2 ... 94 95 96 97 98 99 100 101 102 103 104 ... 187 188

Comentarios a estos manuales

Sin comentarios