6 If you need to compare several single-shot events,
press
Autostore .
Like the Run key, the Autostore key also arms the trigger circuit. When the
trigger conditions are met, the oscilloscope triggers. Pressing the Autostore
key again rearms the trigger circuit without erasing the display. All the data
points are retained on the display in half bright with each trigger allowing you
to easily compare a series of single-shot events.
After you have acquired a single-shot event, pressing a front-panel key,
softkey, or changing a knob can erase the event from the display. If you
press the Stop key, the oscilloscope will recover the event and restore the
oscilloscope settings.
To clear the display, press Erase .
To exit the Autostore mode, press either Run
or Autostore . Notice that RUN replaces STORE in the status line,
indicating that the oscilloscope has exited the Autostore mode.
Operating hint
With display vectors on, the maximum single-shot bandwidth is:
54615B250 MHz for single- and two-channel operation (1 GSa/s, normal
display, display vectors on.)
54616B/16C – 500 MHz for single- and two-channel operation (2 GSa/s,
normal display, display vectors on.)
With display vectors off, the oscilloscopes display the actual captured samples.
Operating Your Oscilloscope
To capture a single event
2–9