Stable trigger, but the waveform is not synchronized with the trigger
Holdoff synchronizes the waveform with the trigger
In Figure 2-4, the holdoff is set to about 25 µs (the duration of the pattern.)
Figure 2-3
Figure 2-4
Operating Your Oscilloscope
To trigger on a complex waveform
2–13