Time Measurements Four additional time measurement choices are available;
+Width, -Width, Rise Time, and Fall Time. Width measurements are made at
the 50% levels, whereas rise time and fall time measurements are made at the
10% to 90% levels.
Previous Menu Returns to the previous softkey menu.
4
Press the Rise Time softkey.
The oscilloscope automatically measures the rise time of the signal and
displays the result on the display.
The oscilloscope makes automatic measurements on the first displayed
event. Figure 2-7 shows how to use delayed sweep to isolate an edge for a
rise time measurement.
Delayed sweep isolates a leading edge for a rise time measurement
Figure 2-7
Operating Your Oscilloscope
To make time measurements automatically
2–18