To verify trigger sensitivity
In this test you verify the trigger sensitivity by applying 100 MHz to the
oscilloscope. The amplitude of the signal is decreased to the specified levels,
then you check to see if the oscilloscope is still triggered. You then repeat
the process at the upper bandwidth limit.
Test limits:
Internal trigger
dc to 100 MHz: 0.5 div or 5.0 mV p-p
100 MHz to 500 MHz: 1 div or 10 mV p-p
External trigger
dc to 100 MHz: <75 mV p-p
100 MHz to 500 MHz: <150 mV p-p
Table 4-7
Equipment Required
Equipment Critical specifications Recommended
Model/Part
Signal generator 100 MHz and 500 MHz sine waves Agilent 8656B
Option 001
Power splitter Outputs differ < 0.15 dB Agilent 11667B
Cable BNC, Qty 3 Agilent 10503A
Adapter N (m) to BNC (f), Qty 3 Agilent 1250-0780
Power meter and
Power sensor
1 to 500 MHz ±3% Agilent 436A and
Agilent 8482A
Service
Verifying Oscilloscope Performance
4-15