Vista de pagina 189
Command Reference
Display Subsystem (:DISPlay)
Chapter 4176
:REMote
Supported All
:DISPlay:REMote ON|OFF|1|0
:DISPlay:REMote?
This command enables or disables the source’s display updating when the signal
generator is remotely controlled.
ON (1) This choice updates the signal generator display so you can see the
settings as the commands are executed, however, this will degrade the
signal generator speed.
OFF (0) This choice turns off the display updating while further optimizing the
signal generator for speed.
*RST N/A
Key Entry
Update in Remote Off On
Remarks The setting enabled by this command is not affected by signal
generator preset or *RST. However, cycling the signal generator power
will reset it to zero.
[:WINDow][:STATe]
Supported All
:DISPlay[:WINDow][:STATe] ON|OFF|1|0
:DISPlay[:WINDow][:STATe]?
This command is used to either blank out (OFF or 0) the display screen or turn it on (ON
or 1).
*RST N/A
Key Entry N/A
Remarks The setting enabled by this command is not affected by *RST. However,
presetting the signal generator or cycling the power will turn the
display on.
Choices ON OFF 1 0
Choices ON OFF 1 0
Vista de pagina 189
1 2 ... 185 186 187 188 189 190 191 192 193 194 195 ... 361 362

Comentarios a estos manuales