Command Reference
Low Frequency Output Subsystem ([:SOURce]:LFOutput)
Chapter 4 265
:FUNCtion[1]:SWEep:TRIGger
Supported PSG-A Series
[:SOURce]:LFOutput:FUNCtion[1]:SWEep:TRIGger BUS|IMMediate|EXTernal|KEY
[:SOURce]:LFOutput:FUNCtion[1]:SWEep:TRIGger?
This command sets the trigger source for the internally generated swept-sine waveform
signal at the LF output.
BUS This choice enables GPIB triggering using the *TRG or GET command or
LAN triggering using the *TRG command.
IMMediate This choice enables immediate triggering of the sweep event.
EXTernal This choice enables the triggering of a sweep event by an externally
applied signal at the TRIGGER IN connector.
KEY This choice enables triggering through front panel interaction by
pressing the
Trigger hardkey.
*RST IMM
Remarks Refer to “:FUNCtion[1]|2:SHAPe” on page 266 for selecting the
waveform type.
Choices BUS IMMediate EXTernal KEY
Key Entry
Bus Free Run Ext Trigger Key