Command Reference
Amplitude Modulation Subsystem ([:SOURce])
Chapter 4 229
:AM[1]|2:TYPE
Supported PSG-A Series
[:SOURce]:AM[1]|2:TYPE LINear|EXPonential
[:SOURce]:AM[1]|2:TYPE?
This command sets the measurement type and unit for the depth of the AM signal.
LINear This choice enables linear depth values in units of percent/volt.
EXPonential This choice enables exponential depth values in units of dB/volt.
*RST LIN
Key Entry
AM Type LIN EXP
Remarks N/A
:AM[1]|2[:DEPTh]:EXPonential
Supported PSG-A Series
[:SOURce]:AM[1]|2[:DEPTh]:EXPonential <val><unit>
[:SOURce]:AM[1]|2[:DEPTh]:EXPonential?
This commands sets the depth of the AM signal in units of dB/volt.
*RST +4.00000000E+001
Range 0.00–40.00DB
Key Entry
AM Depth
Remarks EXPonential must be the current measurement choice for this
command to have any affect. Refer to “:AM[1]|2:TYPE” for setting the
AM measurement mode.
Choices LINear EXPonential