Command Reference
Amplitude Modulation Subsystem ([:SOURce])
Chapter 4232
:AM[:DEPTh]:STEP[:INCRement]
Supported PSG-A Series
[:SOURce]:AM[:DEPTh]:STEP[:INCRement] <num>
[:SOURce]:AM[:DEPTh]:STEP[:INCRement]?
This command sets the depth increment value for the LINear measurement choice.
The variable <num> sets the increment value in units of percent.
*RST
Ν/Α
Range 0.1–100
Key Entry
Incr Set
Remarks Refer to :AM[1]|2:TYPE” on page 229 for setting the AM
measurement choice.
The value set by this command is used with the UP and DOWN choices
for the AM linear depth command. Refer to
“:AM[1]|2[:DEPTh][:LINear]” on page 230 for more information.
The setting enabled by this command is not affected by signal
generator power-on, preset, or *RST.