Command Reference
System Subsystem (:SYSTem)
Chapter 4212
:PRESet:TYPE
Supported All
:SYSTem:PRESet:TYPE NORMal|USER
:SYSTem:PRESet:TYPE?
This command toggles the preset state between factory- and user-defined conditions.
*RST N/A
Key Entry
Preset Normal User
Remarks Refer to “:PRESet[:USER]:SAVE” for saving the USER choice preset
settings.
The setting enabled by this command is not affected by signal
generator power-on, preset, or *RST.
:PRESet[:USER]:SAVE
Supported All
:SYSTem:PRESet[:USER]:SAVE
This command saves your user-defined preset conditions to a state file.
*RST N/A
Range N/A
Key Entry
Save User Preset
Remarks Only one user-defined preset file can be saved. Subsequent saved
user-defined preset files will overwrite the previously saved file.
Choices NORMal USER