Command Reference
Amplitude Modulation Subsystem ([:SOURce])
Chapter 4 221
:AM:INTernal:FREQuency:STEP[:INCRement]
Supported PSG-A Series
[:SOURce]:AM:INTernal:FREQuency:STEP[:INCRement] <num>
[:SOURce]:AM:INTernal:FREQuency:STEP[:INCRement]?
This command sets the step increment for the amplitude modulation internal frequency.
The variable <num> sets the entered value in units of hertz.
*RST
Ν/Α
Range 0.5–1E6
Key Entry
Incr Set
Remarks The value set by this command is used with the UP and DOWN choices
for the AM frequency setting. Refer to
“:AM[1]|2:INTernal[1]|2:FREQuency” on page 226 for more
information.
The setting enabled by this command is not affected by signal
generator power-on, preset, or *RST.