Agilent Technologies PSG Manual de instrucciones Pagina 291

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 362
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 290
Command Reference
Phase Modulation subsystem ([:SOURce])
Chapter 4 277
:PM[1]|2[:DEViation]
Supported PSG-A Series
[:SOURce]:PM[1]|2[:DEViation] <val><unit>|UP|DOWN
[:SOURce]:PM[1]|2[:DEViation]?
This command sets the deviation of the phase modulation.
The variable <unit> will accept RAD (radians), PIRAD (pi-radians), and DEG (degrees);
however, the query will only return values in radians.
*RST
+
0.00000000E
+
000
Range Frequency Normal Bandwidth High Bandwidth
100kHZ
250MHZ 0–10RAD 0–1RAD
> 250
500MHZ 0–5RAD 0–0.5RAD
> 500MHZ
−1
GHZ 0–10RAD 0–1RAD
> 1
2GHZ 0–20RAD 0–2RAD
> 2
3.2GHZ 0–40RAD 0–4RAD
> 3.2
−10
GHZ 0–80RAD 0–8RAD
> 10
20GHZ 0–160RAD 0–16RAD
> 20
40GHZ 0–320RAD 0–32RAD
Choices <val><unit> UP DOWN
Key Entry
ΦM Dev
Remarks If deviation tracking is active, a change to the deviation value on one
path will apply to both.
Refer to “:PM[:DEViation]:STEP[:INCRement]” on page 278 for setting
the value associated with the UP and DOWN choices.