Agilent Technologies PSG Manual de instrucciones Pagina 298

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 362
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 297
Command Reference
Power Subsystem ([:SOURce])
Chapter 4284
:POWer:REFerence
Supported All
[:SOURce]:POWer:REFerence <val><unit>
[:SOURce]:POWer:REFerence?
This command sets the current output power reference.
*RST
+
0.00000000E
+
000
Range
400 to 300DBM
Key Entry
Ampl Ref Set
Remarks The power reference range is affected by power offset.
:POWer:REFerence:STATe
Supported All
[:SOURce]:POWer:REFerence:STATe ON|OFF|1|0
[:SOURce]:POWer:REFerence:STATe?
This command enables or disables the RF output reference.
*RST 0
Key Entry
Ampl Ref Off On
Remarks Once the reference state is ON, all subsequent output power settings
are set relative to the reference value.
Amplitude offsets can be used with the amplitude reference mode.
Choices ON OFF 1 0
Vista de pagina 297
1 2 ... 293 294 295 296 297 298 299 300 301 302 303 ... 361 362

Comentarios a estos manuales

Sin comentarios