Chapter 6: Replacing Assemblies
To remove and replace the front panel bezzle, front panel keyboard, touchscreen, and Autoprobe
board
124
Figure 6-16
13
There is a cable you will need to disconnect from the back of the breakaway board (see Figure
6-17).
Figure 6-17
14
To remove the Autoprobe board, remove the two T10 screws and then slide the board to the
right to disengage the two latches on either side of the board (see Figure 6-18). When
reattaching the Autoprobe board, do not fully tighten the two T10 screws until you have
reattached the front panel keyboard to the oscilloscope and attached it to the Autoprobe
Touchscreen
Push breakaway board
straight out away from
the touchscreen and
then lift up
These are two of the latches you are disengaging
when you pull the board straight away from the
touchscreen
This is the edge that
breaks away from the
front panel keyboard
(they come attached
together)
Disconnect
this cable
Vista de pagina 123