Chapter 4: Testing Performance
Offset Accuracy Test
54
c When the Enter Measurement Info window is displayed, ensure that the V avg function is set
up as follows and then click OK:
Source = Channel 1
Measurement Area = Entire Display
4
Press the Clear Display key on the scope and wait for the #Avgs value (top left corner of screen)
to return to 256. Record the scope's mean V avg reading in the V
zero-error
column (Offset
Accuracy Test tables - both the 50 W table and 1 MW table) of the Performance Test Record
at the end of this chapter.
Notes
• For all scope readings in this procedure, use the mean value in the Measurements display area at the
bottom of the screen.
• If a question mark is displayed in front of any of the values at the bottom of the screen, press the Clear
Display key on the scope, wait for the #Avgs value to return to 256 and then record the scope reading.
The question mark indicates that the oscilloscope could not make a reliable measurement. Therefore,
any results that contain a question mark should be considered suspect and you should follow the
directions above.
Record the
mean reading
Vertical
sensitivity
control for
step 5