Chapter 4: Testing Performance
Analog Bandwidth - Maximum Frequency Check
67
d Set the horizontal scale to 16 ns/div (to display 8 cycles of a 50 MHz waveform).
e Pull down the Setup menu, select Acquisition and then set up the acquisition parameters as
follows:
Analog Memory Depth = Automatic
Analog Sampling rate = Maximum (check the Manual radio button, then click on the field
displaying the value, and then click on the Max button in the dialog box that appears)
Sin(x)/x Interpolation filter enabled
Analog Averaging = Disabled
f Pull down the Measure menu, select Voltage and then select V rms.
Click here and enter 16E-9