Agilent Technologies Ininiium 9000 Especificaciones Pagina 130

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 166
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 129
Chapter 6: Replacing Assemblies
To remove and replace the touch screen controller board
130
To remove and replace the touch screen controller board
Use this procedure to disassemble and reassemble the touch screen controller board. When necessary,
refer to other removal procedures. The graphics in this chapter are representative of the oscilloscope at
the time of this printing. Your unit may look different.
1
Disconnect the power cable and remove the handle and rear cover.
2 Remove the front panel bezzle as described in a previous section.
3 Open up the front chassis by removing the four T15 screws (two are located on the right side
of the oscilloscope and two are located on the left side). See Figure 6-27.
Figure 6-27
4
Pull the front panel chassis away from the main chassis as shown in Figure 6-28.
This is the upper right side
screw that needs removed.
There is also one below it and
then two screws on the
left side of the oscilloscope
in the same relative locations
Vista de pagina 129
1 2 ... 125 126 127 128 129 130 131 132 133 134 135 ... 165 166

Comentarios a estos manuales

Sin comentarios