Chapter 5: Troubleshooting
Power Supply Trouble Isolation
91
Power Supply Trouble Isolation
WARNING SHOCK HAZARD!
Only trained service personnel who are aware of the hazards involved should perform the maintenance
Read the safety summary at the back of this book before proceeding. Failure to observe safety
precautions may result in electric shock
For information on how to replace assemblies, see "Replacing Assemblies," beginning on page 113.
These trouble isolation instructions help isolate the problem to the assembly level when the power system
is not operating. Because of advanced power system protection features, the problem may not be with
the supply itself and, therefore, you will need to use the LEDs located in the upper right hand corner of
the acquisition board to determine the fault. These LEDs can be seen by looking through the rear fan
slots in the upper right hand corner as viewed from the rear of the oscilloscope. After a shutdown occurs,
keep the power cord plugged in so the LEDs remain illuminated. Once you have noted the LED pattern,
you can unplug the instrument.
If the oscilloscope still will not power on after you have followed the required steps below, refer to the
Motherboard Verification section in this chapter for additional diagnosis suggestions.
Look through these
slots and the fans
located behind
them to see the
upper right hand
corner of the
acquisition board
Comentarios a estos manuales