Chapter 6: Replacing Assemblies
To remove and replace the backlight inverter board
129
Figure 6-25
5
Disconnect the cables from the backlight inverter board and remove the two T10 screws (shown
in Figure 6-26). You may need to temporarily remove the touch screen cable (with the blue
connector) in order to remove and replace the inverter board.
Figure 6-26
6
When you replace the inverter board, tighten the screws to 5 in-lbs and be sure to align the
board over the pems.
Main oscilloscope chassis (lying on
its rear panel)
Front chassis after you lift it up
Inverter board
Disconnect these cables
Remove these two T10 screws