Adjustments ESG Family Signal Generators
DCFM Calibration
5-24 Calibration Guide
DCFM Calibration
Description
This test removes all of the dc offsets associated with the FM path while in DCFM mode.
This test uses only the FM1 path to verify the functionality of the circuitry. The resulting
values are stored in the signal generator’s firmware.
Required Test Equipment
HP/Agilent 53132A Option 050 Universal Counter
Procedure
Figure 5-20 DCFM Calibration Setup
1. Connect the equipment as shown above.
2. Preset all of the equipment.
3. Follow the instructions as they appear on the controller’s display.