Performance Tests ESG Family Signal Generators
Support Software
4-4 Calibration Guide
Figure 4-2 The DUT Selection Window
7. In the Select Test Equipment and Tests window (Figure 4-3), create a list of test
equipment that will verify the signal generator’s performance:
a. In the Device Type list (item 6), select a device type that you want to add to the
equipment list.
b. In the Model list (item 7), select the model of the device type that you want to add to
the equipment list.
c. In the Available Test Equipment box (item 8), select the serial number of an
instrument. There may be more than one instrument available for any model. Make
sure that the instrument calibration due date has not passed. If the due date has
passed, use another available instrument. Also, alert your system administrator that
the instrument is due for calibration.
d. Click Add (item 9) to add the instrument to the test equipment list.
e. Continue adding test equipment to the list by repeating these steps.
NOTE You can correct mistakes while you are entering instruments in the test
equipment list:
To delete a single entry in the equipment list, click on the unwanted
selection in the Selected Test Equipment box and then click Remove (item
11).
To delete the entire list of equipment, click Clear (item 12).