ESG Family Signal Generators Adjustments
ALC Modulation Flatness Calibration
Calibration Guide 5-39
ALC Modulation Flatness Calibration
Description
This test calibrates the ALC REF DAC corrections necessary to maintain power flatness in
the ALC open-loop mode of operation. The nominal ALC REF DAC settings are adjusted to
re-align the ALC open-loop power with the desired level. The resulting values are stored as
calibration arrays in the signal generator’s firmware.
Required Test Equipment
• HP/Agilent 438A Power Meter
• HP/Agilent 8482A Power Sensor
Procedure
Figure 5-33 ALC Modulation Flatness Calibration Setup
1. Connect the equipment as shown above.
2. Preset all of the equipment.
3. Follow the instructions as they appear on the controller’s display.