ESG Family Signal Generators Operation Verification
Verification Procedures
Calibration Guide 2-5
Configure the Signal Generator
1. Preset the signal generator to normal (factory-defined) preset conditions.
2. Press
Frequency. Enter 277 using the numeric keypad and press the kHz terminator
softkey.
3. Press Amplitude. Enter 13 using the numeric keypad and press the dBm terminator
softkey.
4. Press Mod On/Off. The MOD OFF annunciator is displayed.
5. Press
RF On/Off. The RF ON annunciator is displayed.
Configure the Power Meter
1. Zero and calibrate the power meter to the sensor’s reference calibration factor.
2. Set the power meter to measure absolute log power (dBm).
Measure the Power Level Accuracy
1. Set the power levels and frequencies for the signal generator (to the maximum
frequency of your signal generator). Refer to the appropriate table in the following list:
Table 2-2, “Power Level Accuracy Setup 1 (Signal Generators without Option 1E6),
on page 2-11
Table 2-3, “Power Level Accuracy Setup 1 (Signal Generators with Option 1E6), on
page 2-13.
2. Set the power sensor’s calibration factor for each frequency.
3. Record the measured power level in the appropriate table (Table 2-2 or Table 2-3) and
compare it to the corresponding limits.
Connect the Test Equipment for Setup 2
Figure 2-3 Power Level Accuracy Equipment Setup 2