ESG Family Signal Generators Operation Verification
Verification Procedures
Calibration Guide 2-11
Table 2-2 Power Level Accuracy Setup 1 (Signal Generators without Option 1E6)
Frequency
Setting
Power Level
Setting
(dBm)
Lower Limit
(dBm)
Measured
Power
(dBm)
Upper Limit
(dBm)
277 kHz +13 12.2 13.8
+7 6.5 7.5
0 −0.5 0.5
−5 −5.5 −4.5
−15 −15.5 −14.5
2.516 MHz +13 12.2 13.8
+7 6.5 7.5
0 −0.5 0.5
−5 −5.5 −4.5
−15 −15.5 −14.5
270.1 MHz +13 12.2 13.8
+7 6.5 7.5
0 −0.5 0.5
−5 −5.5 −4.5
−15 −15.5 −14.5
510.1 MHz +13 12.2 13.8
+7 6.5 7.5
0 −0.5 0.5
−5 −5.5 −4.5
−15 −15.5 −14.5
990.1 MHz +13 12.2 13.8
+7 6.5 7.5
0 −0.5 0.5
−5 −5.5 −4.5
−15 −15.5 −14.5
1350.1 MHz +10 9.2 10.8
+7 6.5 7.5
0 −0.5 0.5
Comentarios a estos manuales