Adjustments ESG Family Signal Generators
AM Gain Calibration
5-40 Calibration Guide
AM Gain Calibration
Description
This test adjusts the gain of the AM circuitry to provide a 10 dB drop for a 1 V input signal.
First a power level is set and a one-volt signal is connected to the input. AM is enabled and
the AM DACs are adjusted for an exact 10 dB drop. The resulting DAC value is stored as a
calibration array.
Required Test Equipment
HP/Agilent 438A Power Meter
HP/Agilent 8482A Power Sensor
HP/Agilent 33120A 15 MHz Function/Arbitrary Waveform Generator
HP/Agilent 3458A Digital Multimeter
Procedure
Figure 5-34 AM Gain Calibration Setup
1. Connect the equipment as shown above.
2. Preset all of the equipment.
3. Follow the instructions as they appear on the controller’s display.