ESG Family Signal Generators Operation Verification
Verification Procedures
Calibration Guide 2-13
Table 2-3 Power Level Accuracy Setup 1 (Signal Generators with Option 1E6)
Frequency
Setting
Power Level
Setting
(dBm)
Lower Limit
(dBm)
Measured
Power
(dBm)
Upper Limit
(dBm)
277 kHz +9 8.2 9.8
+3 2.5 3.5
0 −0.5 0.5
−5 −5.5 −4.5
−15 −15.5 −14.5
2.516 MHz +9 8.2 9.8
+3 2.5 3.5
0 −0.5 0.5
−5 −5.5 −4.5
−15 −15.5 −14.5
270.1 MHz +9 8.2 9.8
+3 2.5 3.5
0 −0.5 0.5
−5 −5.5 −4.5
−15 −15.5 −14.5
510.1 MHz +9 8.2 9.8
+3 2.5 3.5
0 −0.5 0.5
−5 −5.5 −4.5
−15 −15.5 −14.5
990.1 MHz +9 8.2 9.8
+3 2.5 3.5
0 −0.5 0.5
−5 −5.5 −4.5
−15 −15.5 −14.5
1350.1 MHz +6 5.2 6.8
+3 2.5 3.5
0 −0.5 0.5
Comentarios a estos manuales