Adjustments ESG Family Signal Generators
Low Noise Floor Gain Adjust (ESG-APand ESG-DP only)
5-32 Calibration Guide
Low Noise Floor Gain Adjust
(ESG-AP and ESG-DP only)
Description
This adjustment calibrates the gain adjust DAC on the output module for low noise floor
(LNF) mode. The instrument is set up with attenuation applied by way of the burst
modulator. This ensures that the RF chain is not in compression. The gain adjust DAC is
then set for a specific RF output power, as measured with a power meter.
Required Test Equipment
HP/Agilent 438A Power Meter
HP/Agilent 8482A Power Sensor
Procedure
Figure 5-28 LNF Gain Adjust Setup
1. Connect the equipment as shown above.
2. Preset all of the equipment.
3. Follow the instructions as they appear on the controller’s display.