Performance Tests ESG Family Signal Generators
6. Phase Modulation Frequency Response
4-16 Calibration Guide
6. Phase Modulation Frequency Response
This automated test verifies the phase modulation frequency response specifications. The
equipment measures the variations in phase deviations due to changes in the applied
ΦM rate; dc to 100 kHz. The response is relative to a 1 kHz reference at the same phase
deviation, and expressed in dB. Each frequency is tested with the following sequence:
1. A reference signal is measured.
2. The set deviation is measured for each applied rate change.
3. Each deviation measurement is compared to the reference signal deviation.
The comparison is a calculation of the difference in the deviations relative to the reference
rate. The relative value is the ΦM frequency response.
Recommended Equipment
HP/Agilent 8902A Measuring Receiver Option 010
HP/Agilent 8904A Function Generator
Equipment Setup
Figure 4-14 Phase Modulation Frequency Response Test Setup