Service Software ESG Family Signal Generators
ESG Family Support Software Administration
3-10 Calibration Guide
Table 3-1 An Example of Calibration Data for Power Sensors
Figure 3-4 Adding the Equipment Information Using the New Test Equipment
Window
6. Click OK.
NOTE The serial number of the test equipment added will be displayed in the
Equipment field of the Test Equipment dialog box (Figure 3-3).
Removing Test Equipment
Removal of test equipment is accomplished using the Test Equipment dialog box. Refer to
Figure 3-5.
1. Select the Device Type of the test equipment to be removed.
Frequency (MHz)
Calibration
Factor (%)
0.1000 97.6
0.3000 98.9
1.0 99.1
3.0 99.4
Vista de pagina 39