ESG Family Signal Generators Performance Tests
11. Timebase Aging Rate (ESG-AP, ESG-DP or Option 1E5 only)
Calibration Guide 4-25
Equipment Setup
Figure 4-23 Timebase Aging Rate Test Setup
Procedure
1. Preset all instruments and let them warm up for at least one hour.
2. If the oscilloscope does not have a 50 input impedance, connect channel 1 through a
50 feedthrough.
3. On the oscilloscope, adjust the external triggering for a display of the 10 MHz REF
OUTPUT signal from the synthesizer.
a. On Channel 1 of the oscilloscope, set the following:
Display: On
Volts/Division: 500 mV
Input Coupling: dc
Input Impedance: 50 ohms (or use a 50 ohm feedthrough)
b. On Channel 2 of the oscilloscope, set the following:
Display: Off
Input Coupling: dc
Input Impedance: 50 ohms (or use a 50 ohm feedthrough)
c. On the Timebase of the oscilloscope, set the following:
Time/Division: 5 ns
d. On the Trigger of the oscilloscope, set the following:
Trigger Source: CH 2
Trigger Mode: Normal
Trigger Level: 0 V