244 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
11 :DEMO Commands
Query Syntax :DEMO:FUNCtion?
The :DEMO:FUNCtion? query returns the currently selected demo signal
type.
Return Format
<signal><NL>
<signal> ::= {SIN | NOIS | PHAS | RING | SINGl | AM | CLK | GLIT
| BURS | MSO | RUNT | TRAN | RFB | SHOL | LFS | CAN
| LIN | UART | I2C | SPI | I2S | CANL}
See Also "Introduction to :DEMO Commands" on page 241
I2S Off
Signals are internally routed to digital
channels D7 through D9:
D9 — SDATA, TTL level, with
"standard" alignment (internally
routed to digital input).
D8 — SCLK, TTL level, (internally
routed to digital input).
D7 — WS, TTL level, low for left
channel, high for right channel
(internally routed to digital input)
Off
CANLin CAN_L, 250 kbps dominant-low,
~2.8 Vpp, ~1.4 V offset
LIN, 19.2 kbps, ~2.8 Vpp, ~1.4 V
offset
Demo Signal
Function
Demo 1 Terminal Demo 2 Terminal