490 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
24 :SBUS<n> Commands
:SBUS<n>:I2S:TRIGger:RANGe
(see page 860)
Command Syntax
:SBUS<n>:I2S:TRIGger:RANGe <lower>,<upper>
<lower> ::= 32-bit integer in signed decimal, <nondecimal>
or <string>
<upper> ::= 32-bit integer in signed decimal, <nondecimal>,
or <string>
<nondecimal> ::= #Hnn...n where n ::= {0,..,9 | A,..,F}
for hexadecimal
<nondecimal> ::= #Bnn...n where n ::= {0 | 1} for binary
<string> ::= "0xnn...n" where n ::= {0,..,9 | A,..,F} for hexadecimal
The :SBUS<n>:I2S:TRIGger:RANGe command sets the lower and upper
range boundaries used by the INRange, OUTRange, INCReasing, and
DECReasing trigger conditions. You can enter the parameters in any order
— the smaller value becomes the <lower> and the larger value becomes the
<upper>.
Note that for INCReasing and DECReasing, the <lower> and <upper> values
correspond to the "Armed" and "Trigger" softkeys.
Query Syntax
:SBUS<n>:I2S:TRIGger:RANGe?
The :SBUS<n>:I2S:TRIGger:RANGe? query returns the currently set lower
and upper range boundaries.
Return Format
<lower>,<upper><NL>
<lower> ::= 32-bit integer in signed decimal, <nondecimal>
or <string>
<upper> ::= 32-bit integer in signed decimal, <nondecimal>,
or <string>
<nondecimal> ::= #Hnn...n where n ::= {0,..,9 | A,..,F}
for hexadecimal
<nondecimal> ::= #Bnn...n where n ::= {0 | 1} for binary
<string> ::= "0xnn...n" where n ::= {0,..,9 | A,..,F} for hexadecimal
See Also "Introduction to :TRIGger Commands" on page 641
NOTE
The length of the <lower> and <upper> values is determined by the :SBUS<n>:I2S:RWIDth
and :SBUS<n>:I2S:TWIDth commands. When the receiver word size is less than the
transmitter word size, the length is equal to the receiver word size. When the receiver word
size is greater than the transmitter word size, the length is equal to the transmitter word
size.