690 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
28 :TRIGger Commands
:TRIGger:SHOLd:TIME:SETup
(see page 860)
Command Syntax
:TRIGger:SHOLd:TIME:SETup <time>[suffix]
<time> ::= floating-point number in NR3 format
[suffix] ::= {s | ms | us | ns | ps}
The :TRIGger:SHOLd:TIME:SETup command allows triggering on a specific
line of video. The line number limits vary with the standard and mode, as
shown in the following table.
Query Syntax
:TRIGger:SHOLd:TIME:SETup?
The :TRIGger:SHOLd:TIME:SETup? query returns the current TV trigger
line number setting.
Return Format
<time><NL>
<time> ::= floating-point number in NR3 format
See Also "Introduction to :TRIGger Commands" on page 641