Agilent Technologies InfiniiVision 3000 DSO-X 3054A Manual de servicio Pagina 638

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 970
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 637
638 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
27 :TIMebase Commands
:TIMebase:WINDow:RANGe
(see page 860)
Command Syntax
:TIMebase:WINDow:RANGe <range value>
<range value> ::= range value in seconds in NR3 format
The :TIMebase:WINDow:RANGe command sets the full- scale horizontal time
in seconds for the zoomed (delayed) window. The range is 10 times the
current zoomed view window seconds per division setting. The main sweep
range determines the range for this command. The maximum value is one
half of the :TIMebase:RANGe value.
Query Syntax
:TIMebase:WINDow:RANGe?
The :TIMebase:WINDow:RANGe? query returns the current window
timebase range setting.
Return Format
<value><NL>
<value> ::= range value in seconds
See Also "Introduction to :TIMebase Commands" on page 630
":TIMebase:RANGe" on page 633
":TIMebase:POSition" on page 632
":TIMebase:SCALe" on page 635
Vista de pagina 637
1 2 ... 633 634 635 636 637 638 639 640 641 642 643 ... 969 970

Comentarios a estos manuales

Sin comentarios