:TRIGger Commands 28
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 695
:TRIGger:TRANsition:TIME
(see page 860)
Command Syntax
:TRIGger:TRANsition:TIME <time>[suffix]
<time> ::= floating-point number in NR3 format
[suffix] ::= {s | ms | us | ns | ps}
The :TRIGger:TRANsition:TIME command sets the time value for rise/fall
time triggers. You also use the :TRIGger:TRANsition:QUALifier command to
specify whether you are triggering on times greater than or less than this
time value.
Query Syntax
:TRIGger:TRANsition:TIME?
The :TRIGger:TRANsition:TIME? query returns the current rise/fall time
trigger time value.
Return Format
<time><NL>
<time> ::= floating-point number in NR3 format
See Also "Introduction to :TRIGger Commands" on page 641
":TRIGger:TRANsition:QUALifier" on page 692