618 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
26 :SYSTem Commands
:SYSTem:DSP
(see page 860)
Command Syntax
:SYSTem:DSP <string>
<string> ::= quoted ASCII string (up to 75 characters)
The :SYSTem:DSP command writes the quoted string (excluding quotation
marks) to a text box in the center of the display. Use :SYStem:DSP "" to
remotely remove the message from the display. (Two sets of quote marks
without a space between them creates a NULL string.) Press any menu key
to manually remove the message from the display.
See Also "Introduction to :SYSTem Commands" on page 616