:TRIGger Commands 28
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 693
:TRIGger:TRANsition:SLOPe
(see page 860)
Command Syntax
:TRIGger:TRANsition:SLOPe <slope>
<slope> ::= {NEGative | POSitive}
The :TRIGger:TRANsition:SLOPe command specifies a POSitive rising edge
or a NEGative falling edge.
Query Syntax
:TRIGger:TRANsition:SLOPe?
The :TRIGger:TRANsition:SLOPe? query returns the current rise/fall time
trigger slope setting.
Return Format
<slope><NL>
<slope> ::= {NEG | POS}
See Also "Introduction to :TRIGger Commands" on page 641
":TRIGger:MODE" on page 648
":TRIGger:TRANsition:SOURce" on page 694