Agilent Technologies InfiniiVision 3000 DSO-X 3054A Manual de servicio Pagina 466

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 970
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 465
466 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
24 :SBUS<n> Commands
:SBUS<n>:CAN:SIGNal:BAUDrate
(see page 860)
Command Syntax
:SBUS<n>:CAN:SIGNal:BAUDrate <baudrate>
<baudrate> ::= integer from 10000 to 1000000 in 100 b/s increments
The :SBUS<n>:CAN:SIGNal:BAUDrate command sets the standard baud rate
of the CAN signal from 10 kb/s to 1 Mb/s in 100 b/s increments. If you
enter a baud rate that is not divisible by 100 b/s, the baud rate is set to
the nearest baud rate divisible by 100 b/s.
If the baud rate you select does not match the system baud rate, false
triggers may occur.
Query Syntax
:SBUS<n>:CAN:SIGNal:BAUDrate?
The :SBUS<n>:CAN:SIGNal:BAUDrate? query returns the current CAN baud
rate setting.
Return Format
<baudrate><NL>
<baudrate> ::= integer from 10000 to 1000000 in 100 b/s increments
See Also "Introduction to :TRIGger Commands" on page 641
":SBUS<n>:MODE" on page 457
":SBUS<n>:CAN:TRIGger" on page 469
":SBUS<n>:CAN:SIGNal:DEFinition" on page 467
":SBUS<n>:CAN:SOURce" on page 468
Vista de pagina 465
1 2 ... 461 462 463 464 465 466 467 468 469 470 471 ... 969 970

Comentarios a estos manuales

Sin comentarios