Agilent Technologies InfiniiVision 3000 DSO-X 3054A Manual de servicio Pagina 866

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 970
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 865
866 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
36 More About Oscilloscope Commands
:<subsystem>:<function><separator><data>;<function><separator><data><ter
minator>
For example:
myScope.WriteString ":TIMebase:RANGe 0.5;POSition 0"
Example 2:
Program
Message
Terminator Sets
Parser Back to
Root
myScope.WriteString ":TIMebase:REFerence CENTer;POSition 0.00001"
or
myScope.WriteString ":TIMebase:REFerence CENTer"
myScope.WriteString ":TIMebase:POSition 0.00001"
A second way to send these commands is by placing TIMebase: before the
POSition command as shown in the second part of example 2. The space
after POSition is required.
Example 3:
Selecting
Multiple
Subsystems
You can send multiple program commands and program queries for
different subsystems on the same line by separating each command with a
semicolon. The colon following the semicolon enables you to enter a new
subsystem. For example:
<program mnemonic><data>;:<program mnemonic><data><terminator>
For example:
myScope.WriteString ":TIMebase:REFerence CENTer;:DISPlay:VECTors ON"
Multiple commands may be any combination of compound and simple
commands.
NOTE
The colon between TIMebase and RANGe is necessary because TIMebase:RANGe is a
compound command. The semicolon between the RANGe command and the POSition
command is the required program message unit separator. The POSition command does not
need TIMebase preceding it because the TIMebase:RANGe command sets the parser to the
TIMebase node in the tree.
NOTE
In the first line of example 2, the subsystem selector is implied for the POSition command in
the compound command. The POSition command must be in the same program message as
the REFerence command because the program message terminator places the parser back
at the root of the command tree.
NOTE
The leading colon before DISPlay:VECTors ON tells the parser to go back to the root of the
command tree. The parser can then see the DISPlay:VECTors ON command. The space
between REFerence and CENter is required; so is the space between VECTors and ON.
Vista de pagina 865
1 2 ... 861 862 863 864 865 866 867 868 869 870 871 ... 969 970

Comentarios a estos manuales

Sin comentarios