796 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
32 Obsolete and Discontinued Commands
:MEASure:TMIN
(see page 860)
Command Syntax
:MEASure:TMIN [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n> ::= 1 to (# analog channels) in NR1 format
The :MEASure:TMIN command installs a screen measurement and starts an
X-at- Min-Y measurement on the selected waveform. If the optional source
is specified, the current source is modified.
Query Syntax
:MEASure:TMIN? [<source>]
The :MEASure:TMIN? query returns the horizontal axis value at which the
minimum vertical value occurs on the current source. If the optional
source is specified, the current source is modified. If all channels are off,
the query returns 9.9E+37.
Return Format
<value><NL>
<value> ::= time at minimum in NR3 format
See Also "Introduction to :MEASure Commands" on page 330
":MEASure:TMAX" on page 795
":MEASure:XMAX" on page 382
":MEASure:XMIN" on page 383
NOTE
The :MEASure:TMIN command is obsolete and is provided for backward compatibility to
previous oscilloscopes. Use the :MEASure:XMIN command (see page 383) instead.