654 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
28 :TRIGger Commands
:TRIGger:EBURst:SLOPe
(see page 860)
Command Syntax
:TRIGger:EBURst:SLOPe <slope>
<slope> ::= {NEGative | POSitive}
The :TRIGger:EBURst:SLOPe command specifies whether the rising edge
(POSitive) or falling edge (NEGative) of the Nth edge in a burst will
generate a trigger.
Query Syntax
:TRIGger:EBURst:SLOPe?
The :TRIGger:EBURst:SLOPe? query returns the current Nth edge in a
burst slope.
Return Format
<slope><NL>
<slope> ::= {NEG | POS}
See Also "Introduction to :TRIGger Commands" on page 641
":TRIGger:EBURst:IDLE" on page 653
":TRIGger:EBURst:COUNt" on page 652