:FUNCtion Commands 15
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 277
:FUNCtion:OFFSet
(see page 860)
Command Syntax
:FUNCtion:OFFSet <offset>
<offset> ::= the value at center screen in NR3 format.
The :FUNCtion:OFFSet command sets the voltage or vertical value
represented at center screen for the selected function. The range of legal
values is generally +/- 10 times the current scale of the selected function,
but will vary by function. If you set the offset to a value outside of the
legal range, the offset value is automatically set to the nearest legal value.
Query Syntax
:FUNCtion:OFFSet?
The :FUNCtion:OFFSet? query outputs the current offset value for the
selected function.
Return Format
<offset><NL>
<offset> ::= the value at center screen in NR3 format.
See Also "Introduction to :FUNCtion Commands" on page 271
":FUNCtion:RANGe" on page 279
":FUNCtion:REFerence" on page 280
":FUNCtion:SCALe" on page 281
NOTE
The :FUNCtion:OFFset command is equivalent to the :FUNCtion:REFerence command.