668 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
28 :TRIGger Commands
:TRIGger:GLITch:QUALifier
(see page 860)
Command Syntax
:TRIGger:GLITch:QUALifier <operator>
<operator> ::= {GREaterthan | LESSthan | RANGe}
This command sets the mode of operation of the glitch pulse width trigger.
The oscilloscope can trigger on a pulse width that is greater than a time
value, less than a time value, or within a range of time values.
Query Syntax
:TRIGger:GLITch:QUALifier?
The :TRIGger:GLITch:QUALifier? query returns the glitch pulse width
qualifier.
Return Format
<operator><NL>
<operator> ::= {GRE | LESS | RANG}
See Also "Introduction to :TRIGger Commands" on page 641
":TRIGger:GLITch:SOURce" on page 670
":TRIGger:MODE" on page 648