1100 LC Modules Control Manual - 152 - 03-914947-30:Rev.6
Maintenance of the Agilent LC 1100 Quaternary pump
(G1311)
Click on the button to display the Agilent Technologies
1100 Quaternary Pump (G1311) diagnostic screen. It contains
four sub-sections: Counters, Logbook, Tests and Actions. Click
on the required sub-section to display the corresponding screen.
Counters Window
This window monitors the usage of several components of the
quaternary pump module which is schematized at the top of the
counters tab.
Two different counters tables are available: one for the pump
components and another for the valves components. Click on the
graphic to select either the pump or valve for display. A box will
be highlighted around the displayed component.
The General and Miscellaneous counters are identical for the
pump and MCGV counters tables. The Valve counters are
displayed only when selecting the MCGV in the module schema.
Select the pump in the module schema. The following counters
limits, actual and accumulated values are displayed and the
pump is bordered with a blue rectangle: