1100 LC Modules Control Manual - 170 - 03-914947-30:Rev.6
For analyzing the test results, refer to the detector operating
manual.
Intensity Test
Perform the Intensity test to evaluate the performance of the
lamps and optics. UV and VIS lamps Intensity are measured
over the detector wavelengths range.
The test should be done with water in the flow cell.
When starting the test, the 1 nm slit is moved into the light path
and the gain is set to zero. Four spectral ranges are used to
evaluate the intensity spectrum. Results are displayed on screen
and also listed in the maintenance audit trail: