
1100 LC Modules Control Manual - 175 - 03-914947-30:Rev.6
During the test, a theoretical analog signal of approximately 52
mV with a square wave of 10µV and a frequency of 1 cycle / 48
seconds is applied to the digital-analog-converter (DAC).
The DAC signal must be plotted on an integrator at least 5
minutes in order to measure the amplitude and the peak-to-peak
noise of the signal. The results are not directly provided by the
Instrument Maintenance Software but by the integrator.
The following performed stages are displayed on screen and
also in the maintenance audit trail:
For analyzing the test results, refer to the detector operating
manual.
Comentarios a estos manuales