1100 LC Modules Control Manual - 163 - 03-914947-30:Rev.6
Pump pistons replaced: no outcome on counters.
Purge valve frit replaced: no outcome on counters
Solvent inlet filter replaced: no outcome on counters
AIV (Active Inlet Valve): Actual Inlet valve switches counters can
be reset.
Outlet valve: Actual Inlet valve switches counters can be reset.
Pump head assembly: no outcome on counters
MCGV (Multi Channel gradient Valve) replaced: MCGV switches
counters from A to D can be reset.
Once an action is performed, it is automatically listed in the
maintenance audit trail.
Maintenance of the Agilent LC 1100 Diode Array Detector
(G1315) and Multiple Wavelength Detector (G1365)
Click on the button to display the Agilent Technologies
1100 Diode Array Detector (G1315) and Multiple Wavelength
detector (G1365) diagnostic screen. It contains four sub-
sections: Counters, Logbook, Tests and Actions. Click on the
required sub-section to display the corresponding screen.
Counters Window
This window monitors the usage of several components of the
detector module which is schematized at the top of the counters
tab.
The following counter limits, actual and accumulated values are
then displayed: