1100 LC Modules Control Manual - 171 - 03-914947-30:Rev.6
For analyzing the test results, refer to the detector operating
manual.
Holmium Test
Perform the Holmium test to verify wavelength accuracy. Three
characteristic absorbance maxima of the holmium oxide filter are
measured.
The test should be done with water in the flow cell.
When starting the test, the 1 nm slit is moved into the light path.
Measured maxima are displayed on screen and also listed in the
maintenance audit trail: