1100 LC Modules Control Manual - 173 - 03-914947-30:Rev.6
For analyzing the test results, refer to the detector operating
manual.
Filter Test
Perform the Filter test to determine the if the filter is clean.
During the test, the filter motor is moved in specific positions and
the resulting absorbance of the holmium oxide filter is monitored.
During the filter movement, the absorbance maxima are
measured and then graphically displayed at screen and in the
maintenance audit trail: