1100 LC Modules Control Manual - 89 - 03-914947-30:Rev.6
Photo Multiplier Tube Parameter Group
Select the value of the Gain parameter (amplification factor).
Lamps Parameter Group
Check Require Flash Lamp if necessary.
Flash Lamp Economy Mode decreases the frequency of the lamp
flash.
Automatic Lamp Stop automatically stops the detector’s lamp at
the end of the run.
Flash Lamp Always On enables permanent flash mode.
Lamp Energy Reference On configures the signal as a reference.
Autozero At Prerun performs an autozero of the detector before
the run.
Control Section for the Agilent LC 1100 Refractive Index
Detector (G1362)
Click on the button to display the Agilent Technologies
1100 Retention Index Detector (G1362) control method section.
It contains three sub-sections: Settings, Analogs and Time table.
Click on the required sub-section to define the corresponding
parameters that will be used during the chromatographic run.